Q. How do I know if isoelectric focusing is proceeding in the IPG strip?
A. As isoelectric focusing proceeds, the bromophenol blue tracking dye migrates toward the anode. Note that the dye front leaves the IPG strip well before focusing is complete, so clearing of the dye is no indication that the sample is focused. If the dye does not migrate, no current is flowing through the IPG strip. Check all connections and the efficiency of rehydration of the IPG strip.
Click here for more information on our Immobiline DryStrips.
|