ImageQuant™ LAS 4000 Biomolecular Imager

  • Exceptional sensitivity for quantitation of biomolecules labeled with chemiluminescent probes, and expandable to multifluorescent UV, visible, and IR imaging. Allows gel documentation of Coomassie™ and silver-stained samples.
  • Accurate quantitation of closely spaced bands in complex protein samples for gels and films up to 21 x 14 cm (25 x 25 cm using the optional lens). Low noise for longer exposure times and lower background, important for precise quantitation of very weak signals.
  • Wide linear dynamic range for precise quantitation of faint and dark bands over four orders of magnitude in the same image, avoiding time consuming multiple film exposures.
  • Distortion, dark frame, and flat frame corrections are applied to each imaging mode for optimal precision and uniform quantitation.
  • Sensitive detection of less than 40 pg of protein on a Western blot. Sensitivity may be increased by "binning" of up to 8 x 8 pixels and smoothing.
  • Chemiluminescent and colorimetric signals can be captured in the same image without changing the lens. Optional multifluorescence detection extends functionality to the capture of fluorophores such as Cy™2, Cy3, and Cy5 and Deep Purple™ Total Protein Stain.
Page down for more information


ImageQuant™ LAS 4000 Biomolecular Imager

Technical Information


Click here to view this image in high resolution
ImageQuant LAS 4000 is a fully featured digital imaging system for sensitive, quantitative imaging of gels, blots, and colonies.

ImageQuant LAS 4000 is designed to quantitate the highly sensitive detection of proteins using Amersham™ ECL™ Western blotting reagents. A free combination of light sources and filters can be incorporated into the imager according to your needs. ImageQuant LAS 4010 is a fully equipped imager that, in addition, performs ultraviolet (UV) and visible (RGB) fluorescence imaging. Both systems can be upgraded to perform infrared (IR) fluorescence imaging.

UV transillumination for documentation of EtBr-stained gels is standard in both systems, and white light transillumination for documentation of Coomassie blue- and silver-stained gels is standard on ImageQuant LAS 4010.

The system consists of an intelligent dark box, camera, lens, and automated four-position filter wheel with a standard filter for EtBr and open positions for the filters delivered with the optional fluorescent light sources. Focusing, filters, illuminators, and exposure time are remotely controlled by a computer. Simple operation allows adjustment without having to open the cabinet. The intelligent light box provides automation and user safety. UV light sources are housed inside the cabinet, and can only be switched on when the door is closed.

A bright, wide aperture FUJINON™ F0.85 lens, specially developed for chemiluminescence imaging, projects sharp images onto a uniquely patterned CCD chip. By diagonally aligning octogonal pixels, the gap between pixels in the horizontal and vertical planes is reduced. This pattern allows an effective image resolution of 6.3 Mpixels.

The user-friendly control software has several capture modes for achieving optimal sensitivity and dynamic range including increment, repetition, and program modes. For image analysis, advanced and flexible ImageQuant TL software is available separately and comprises a dedicated suite of analysis applications to provide high levels of automation and accuracy in analysis of gels, blots, and colony counting.

TECHNICAL SPECIFICATIONS
CCD model Fujifilm Super CCD Area Type chip (15.6 x 23.4 mm) 
Lens model FUJINON™ Lens F0.85 43 mm 
Cooling Two-stage thermoelectric module with air circulation 
Cooling temperature -25°C; max -30°C (at room temperatures below 28°C) 
Dynamic range 16-bit, 4 orders of magnitude 
Chip resolution 2048 x 1472, 3.2 Mpixels 
Image resolution Maximum 3072 x 2048, 6.3 Mpixels 
Pixel size Approximately 11 μm 
Focusing and diaphragm Automatic, remote operation 
Capture modes Automatic, manual (normal/incremental/repetitive/program) 
Exposure time Automatic, manual (1/100 s to 30 h) 
Pixel correction Dark frame correction, flat frame correction, distortion correction 
Image quality correction Binning, smoothing 
Binning modes 1 x 1, 2 x 2, 4 x 4, 8 x 8  
Image Size Up to 12 MB (.GEL and .TIF) 
Sample size  21 x 14 cm (25 x 25 cm with SIGMA wide view lens) 
Interface USB 2.0 
Dimensions (W x H x D) 510 x 900 x 480 mm 
Weight 60 kg 
Line frequency 50 / 60 Hz 
Operating temperature 15°C to 28°C 
Operating humidity 30% to 70% (no condensation) 
Supply voltage 100 to 240 V 
Power consumption Approx. 0.3 kVA 