| ÄKTA™ oligopilot™ plus is a flexible and fully automated DNA/RNA oligonucleotide synthesizer designed for use in research, process development laboratories, and production. ÄKTA oligopilot plus is designed to meet the needs of new synthesis chemistries like RNAi and enables cost-efficient and high-quality synthesis. ÄKTA oligopilot plus is a compact, pump-driven system employing flow-through reactor technology, which results in significantly reduced reagent consumption, enables exact control of reaction rates and contact times, and allows smooth scale up. ÄKTA oligopilot plus features column re-circulation, an important factor when performing RNA synthesis. ÄKTA oligopilot plus is available in two different configurations: ÄKTAoligopilot plus 10 for synthesis in the 1-50 µmol range and ÄKTAoligopilot plus 100 for synthesis in the 50 µmol-9 mmol range. | |
Features and benefits
| |
| |
| Chemistry | Synthesis of DNA, RNA, phosphodiester, and phosphorothioat oligonucleotides using β-cyanoethyl, phosphoramidites. |
| Synthesis scales | 1 µmol to 9 mmol depending on degree of substitution of solid support used. |
| Coupling efficiency | > 99% for DNA >98% for RNA |
| Cycle time | 4.5 to 25 min for standard DNA |
| Reagent stability in instrument | At least 2 wk |
| Controlled by UNICORN software on an external computer. The software is menu and mouse driven and allows method creation, synthesis system control, monitoring, and evaluation. UNICORN software is supplied on a CD. System-specific Strategy and Template methods are supplied on a separate CD. |
| Dimensions: (W x H x D) | 450 x 610 x 480 mm |
| Weight | 63 kg |
| No. of column reactors | Up to seven |
| No of monomer ports | Eight fitted as standard. Option to use additional monomer inlets up to a total of 12. |
| Reagent delivery and washing | Active pumping with a flow rate range of 0.01 to 100 ml ÄKTA oligopilot plus 10: 2 x 10 ml/min ÄKTA oligopilot plus 100: 2 x 100 ml/min |
| Monitoring and control system | Conductivity monitoring. UV-vis monitoring of DMTr. Automatic calcuation of coupling efficiency. Pause capability if synthesis fails. Pressure. |
| Protective system | Argon or Nitrogen atmosphere |
| Delivery system | Eight low dead-volume motorized valves |
| Control system | UNICORN software on an external computer |
| Waste diversion system | Separates chlorinated from nonchlorinated hydrocarbons. Also possible to separate six other solvents, e.g monomer solutions. |
| Methods | Optimized, standard methods for synthesis in the range 1 μmol to 9 mmol. Customize methods by creating new methods or changing existing standard methods. |
| Warranty | One-year parts and labor |
Oligosynt prepacked disposable columns for the synthesis of long oligonucleotides pdf
Primer Support 200 for therapeutic oligonucleotide production pdf
| Get the training you need in chromatography, process development, column packing, filtration or cell culture with Wave™. Fast Trak Training & Education has over 20 years of experience that provide you with the skills and knowledge for your industrial bioprocessing. |